A phantom based evaluation on the effects of patient breathing motion on Stereotactic Body Radiotherapy treatment volumes

Aim: The aim of the study was to design an upper body phantom to mimic the movement of the lesion inside the lungs during a breathing cycle. Phantom design included an assessment of the motion observed for lung lesions, identification of suitable phantom materials as well as design of a motorized ar...

Full description

Saved in:
Bibliographic Details
Main Author: Coetzee, Nicolene
Other Authors: Burger, Hester
Format: Thesis
Language:English
Published: Division of Radiology 2020
Subjects:
Tags: Add Tag
No Tags, Be the first to tag this record!
_version_ 1867613169050976256
access_status_str Open Access
author Coetzee, Nicolene
author2 Burger, Hester
author_browse Burger, Hester
Coetzee, Nicolene
author_facet Burger, Hester
Coetzee, Nicolene
author_sort Coetzee, Nicolene
collection Thesis
description Aim: The aim of the study was to design an upper body phantom to mimic the movement of the lesion inside the lungs during a breathing cycle. Phantom design included an assessment of the motion observed for lung lesions, identification of suitable phantom materials as well as design of a motorized arm to mimic the movements observed inside the lung area of the phantom. Introduction: Expansion margins are added to clinical target volumes contoured by Oncologists in order to safeguard against under- or over-treatment of the target volume. They are designed to account for errors during setup, inaccuracies on the linear accelerator, and movement of targets inside the patient. If the margins are too small, there is a risk that the lesion/target may not receive the necessary dose, due to being partially missed. On the other hand, if the margins are too wide, the lesion will be covered, but normal tissue may receive unnecessary dose, resulting in additional side effects to the patient. Assessment of the impact of these margins is not possible in a static phantom and the availability of a low-cost motorized phantom would assist in the validation of these margins. Method: Previously treated patients' 4D CT scanning data were used to quantify the amount of movement seen for lesions within the lung. A phantom was then designed and built in an attempt to mimic both patient anatomy and movement. Materials were identified to replicate anatomical shape and densities of various organs in the thorax, as seen on CT scan data. Two treatment planning systems (Monaco, (Elekta) and Eclipse (Varian)) were used to determine the dosimetric characteristics of the materials. This was compared to actual dose as delivered by a linear accelerator (Elekta Synergy). Results: Paths were calculated from the breathing cycles during the 4D-CT scan sets and templates designed to mimic these movements. A thorax phantom was built with the appropriate materials suitable and matched densities to replicate a human thorax. Comparing transmission for these materials on a linear accelerator for 6MV and 10MV energy, the deviation from planned versus measured dose varied between 1.67% to 3.32% and 0.45% to 2.30%, respectively for the silicon material and between 0.77% to 3.22% and 0.17% to 2.57% for the 3D printed bone for 6MV and 10MV. iv Conclusion: The measurements done on the linear accelerator matched closely with the calculated values on the treatment planning system for transmission through the materials in the customised phantom. Various proposals were put forward to mimic the movement of the targets within the lung regions. However, it was not possible to manufacture a mechanically based working model due to the small movements observed (<5mm). It is recommended that a robotic solution be investigated as alternative to mimic these small movements.
format Thesis
id oai:open.uct.ac.za:11427/32181
institution University of Cape Town (South Africa)
language eng
last_indexed 2026-06-10T12:31:52.071Z
license_str Not specified — see source repository
provenance_str_mv Harvested via OAI-PMH from UCTD — University of Cape Town Open Access Repository
publishDate 2020
publishDateRange 2020
publishDateSort 2020
publisher Division of Radiology
publisherStr Division of Radiology
record_format dspace
source_str UCTD — University of Cape Town Open Access Repository
spelling oai:open.uct.ac.za:11427/32181 A phantom based evaluation on the effects of patient breathing motion on Stereotactic Body Radiotherapy treatment volumes Coetzee, Nicolene Burger, Hester Joubert Nanette Medicine Aim: The aim of the study was to design an upper body phantom to mimic the movement of the lesion inside the lungs during a breathing cycle. Phantom design included an assessment of the motion observed for lung lesions, identification of suitable phantom materials as well as design of a motorized arm to mimic the movements observed inside the lung area of the phantom. Introduction: Expansion margins are added to clinical target volumes contoured by Oncologists in order to safeguard against under- or over-treatment of the target volume. They are designed to account for errors during setup, inaccuracies on the linear accelerator, and movement of targets inside the patient. If the margins are too small, there is a risk that the lesion/target may not receive the necessary dose, due to being partially missed. On the other hand, if the margins are too wide, the lesion will be covered, but normal tissue may receive unnecessary dose, resulting in additional side effects to the patient. Assessment of the impact of these margins is not possible in a static phantom and the availability of a low-cost motorized phantom would assist in the validation of these margins. Method: Previously treated patients' 4D CT scanning data were used to quantify the amount of movement seen for lesions within the lung. A phantom was then designed and built in an attempt to mimic both patient anatomy and movement. Materials were identified to replicate anatomical shape and densities of various organs in the thorax, as seen on CT scan data. Two treatment planning systems (Monaco, (Elekta) and Eclipse (Varian)) were used to determine the dosimetric characteristics of the materials. This was compared to actual dose as delivered by a linear accelerator (Elekta Synergy). Results: Paths were calculated from the breathing cycles during the 4D-CT scan sets and templates designed to mimic these movements. A thorax phantom was built with the appropriate materials suitable and matched densities to replicate a human thorax. Comparing transmission for these materials on a linear accelerator for 6MV and 10MV energy, the deviation from planned versus measured dose varied between 1.67% to 3.32% and 0.45% to 2.30%, respectively for the silicon material and between 0.77% to 3.22% and 0.17% to 2.57% for the 3D printed bone for 6MV and 10MV. iv Conclusion: The measurements done on the linear accelerator matched closely with the calculated values on the treatment planning system for transmission through the materials in the customised phantom. Various proposals were put forward to mimic the movement of the targets within the lung regions. However, it was not possible to manufacture a mechanically based working model due to the small movements observed (<5mm). It is recommended that a robotic solution be investigated as alternative to mimic these small movements. 2020-09-09T09:12:51Z 2020-09-09T09:12:51Z 2020 2020-09-09T06:34:30Z Master Thesis Masters MSc http://hdl.handle.net/11427/32181 eng application/pdf Division of Radiology Faculty of Health Sciences
spellingShingle Medicine
Coetzee, Nicolene
A phantom based evaluation on the effects of patient breathing motion on Stereotactic Body Radiotherapy treatment volumes
thesis_degree_str Master's
title A phantom based evaluation on the effects of patient breathing motion on Stereotactic Body Radiotherapy treatment volumes
title_full A phantom based evaluation on the effects of patient breathing motion on Stereotactic Body Radiotherapy treatment volumes
title_fullStr A phantom based evaluation on the effects of patient breathing motion on Stereotactic Body Radiotherapy treatment volumes
title_full_unstemmed A phantom based evaluation on the effects of patient breathing motion on Stereotactic Body Radiotherapy treatment volumes
title_short A phantom based evaluation on the effects of patient breathing motion on Stereotactic Body Radiotherapy treatment volumes
title_sort phantom based evaluation on the effects of patient breathing motion on stereotactic body radiotherapy treatment volumes
topic Medicine
url http://hdl.handle.net/11427/32181
work_keys_str_mv AT coetzeenicolene aphantombasedevaluationontheeffectsofpatientbreathingmotiononstereotacticbodyradiotherapytreatmentvolumes
AT coetzeenicolene phantombasedevaluationontheeffectsofpatientbreathingmotiononstereotacticbodyradiotherapytreatmentvolumes