The effect of device position and use of transparent covers on the irradiance distribution of LED phototherapy devices

Background Effective phototherapy reduces neonatal jaundice and its complications. Irradiance increases as the distance of the light source decreases from a single phototherapy light. There are limited studies of the effect of distance and positional changes on different LED light designs on achievi...

Full description

Saved in:
Bibliographic Details
Main Author: Ismail, Mugammad Taib
Other Authors: Horn, Alan
Format: Thesis
Language:English
Published: Department of Paediatrics and Child Health 2021
Subjects:
Tags: Add Tag
No Tags, Be the first to tag this record!
_version_ 1867613234595364864
access_status_str Open Access
author Ismail, Mugammad Taib
author2 Horn, Alan
author_browse Horn, Alan
Ismail, Mugammad Taib
author_facet Horn, Alan
Ismail, Mugammad Taib
author_sort Ismail, Mugammad Taib
collection Thesis
description Background Effective phototherapy reduces neonatal jaundice and its complications. Irradiance increases as the distance of the light source decreases from a single phototherapy light. There are limited studies of the effect of distance and positional changes on different LED light designs on achieving effective phototherapy. Objectives To describe and compare the effect of distance, angle and plastic barriers on three different LED lights of different design. Methods Comparisons were made using a Servolite LED light, a General Electric (GE) Lullaby and a Ningbo David LED phototherapy light. Measurements were done according to methods described by the International Electrotechnical Communission (IEC). The effective irradiated area was measured on a grid measuring 60 x 30 cm subdivided into 5 x 5 cm squares. Measurements were done for the following scenarios: light placed at the manufacturers' recommended distance, 20 cm closer, 20 cm further, at an angle, through clear plastic and through scuffed perspex. Results When the lights were placed closer to the irradiated surface than the manufacturers' recommendations, the maximum irradiance increased, but the median irradiance and uniformity ratio decreased. When the lights were angled at 45 the median irradiance was decreased. A decrease in the median irradiance was also seen when phototherapy lights passed through scuffed plastic and food grade plastic. Conclusion Our study demonstrated that placing LED lights closer than the manufacturers recommendations, the use of transparent barriers and the use of lights at an angle, compromised phototherapy irradiance and distribution. Only the GE light met IEC standards.
format Thesis
id oai:open.uct.ac.za:11427/33717
institution University of Cape Town (South Africa)
language eng
last_indexed 2026-06-10T12:32:54.720Z
license_str Not specified — see source repository
provenance_str_mv Harvested via OAI-PMH from UCTD — University of Cape Town Open Access Repository
publishDate 2021
publishDateRange 2021
publishDateSort 2021
publisher Department of Paediatrics and Child Health
publisherStr Department of Paediatrics and Child Health
record_format dspace
source_str UCTD — University of Cape Town Open Access Repository
spelling oai:open.uct.ac.za:11427/33717 The effect of device position and use of transparent covers on the irradiance distribution of LED phototherapy devices Ismail, Mugammad Taib Horn, Alan paediatrics and child health Background Effective phototherapy reduces neonatal jaundice and its complications. Irradiance increases as the distance of the light source decreases from a single phototherapy light. There are limited studies of the effect of distance and positional changes on different LED light designs on achieving effective phototherapy. Objectives To describe and compare the effect of distance, angle and plastic barriers on three different LED lights of different design. Methods Comparisons were made using a Servolite LED light, a General Electric (GE) Lullaby and a Ningbo David LED phototherapy light. Measurements were done according to methods described by the International Electrotechnical Communission (IEC). The effective irradiated area was measured on a grid measuring 60 x 30 cm subdivided into 5 x 5 cm squares. Measurements were done for the following scenarios: light placed at the manufacturers' recommended distance, 20 cm closer, 20 cm further, at an angle, through clear plastic and through scuffed perspex. Results When the lights were placed closer to the irradiated surface than the manufacturers' recommendations, the maximum irradiance increased, but the median irradiance and uniformity ratio decreased. When the lights were angled at 45 the median irradiance was decreased. A decrease in the median irradiance was also seen when phototherapy lights passed through scuffed plastic and food grade plastic. Conclusion Our study demonstrated that placing LED lights closer than the manufacturers recommendations, the use of transparent barriers and the use of lights at an angle, compromised phototherapy irradiance and distribution. Only the GE light met IEC standards. 2021-08-06T08:56:42Z 2021-08-06T08:56:42Z 2021 2021-08-06T08:55:50Z Master Thesis Masters MMed http://hdl.handle.net/11427/33717 eng application/pdf Department of Paediatrics and Child Health Faculty of Health Sciences
spellingShingle paediatrics and child health
Ismail, Mugammad Taib
The effect of device position and use of transparent covers on the irradiance distribution of LED phototherapy devices
thesis_degree_str Master's
title The effect of device position and use of transparent covers on the irradiance distribution of LED phototherapy devices
title_full The effect of device position and use of transparent covers on the irradiance distribution of LED phototherapy devices
title_fullStr The effect of device position and use of transparent covers on the irradiance distribution of LED phototherapy devices
title_full_unstemmed The effect of device position and use of transparent covers on the irradiance distribution of LED phototherapy devices
title_short The effect of device position and use of transparent covers on the irradiance distribution of LED phototherapy devices
title_sort effect of device position and use of transparent covers on the irradiance distribution of led phototherapy devices
topic paediatrics and child health
url http://hdl.handle.net/11427/33717
work_keys_str_mv AT ismailmugammadtaib theeffectofdevicepositionanduseoftransparentcoversontheirradiancedistributionofledphototherapydevices
AT ismailmugammadtaib effectofdevicepositionanduseoftransparentcoversontheirradiancedistributionofledphototherapydevices