Design and Development of an Open-Source ADL-Compliant Prosthetic Arm for Transradial Amputees

Transradial amputation is traumatic – leading to the amputee having a limited ability to perform activities of daily living (ADLs). Below-elbow prostheses are prescribed. The high cost associated with prostheses results in many amputees in low-to-middleincome countries relying on government subsidis...

Full description

Saved in:
Bibliographic Details
Main Author: Timm, Lara
Other Authors: Sivarasu, Sudesh
Format: Thesis
Language:English
Published: Department of Human Biology 2022
Subjects:
Tags: Add Tag
No Tags, Be the first to tag this record!
_version_ 1867613210633306112
access_status_str Open Access
author Timm, Lara
author2 Sivarasu, Sudesh
author_browse Sivarasu, Sudesh
Timm, Lara
author_facet Sivarasu, Sudesh
Timm, Lara
author_sort Timm, Lara
collection Thesis
description Transradial amputation is traumatic – leading to the amputee having a limited ability to perform activities of daily living (ADLs). Below-elbow prostheses are prescribed. The high cost associated with prostheses results in many amputees in low-to-middleincome countries relying on government subsidised devices, which are cosmetic rather than functional, or none at all. Open-source prostheses have the potential to increase the accessibility of functional prosthetic arms, but at present are not optimised to assist the dominant hand in performing bimanual ADLs. The aim of this study is thus to design and experimentally validate an open-source prosthetic arm that is functionally optimised for the performance of ADLs in the unilateral transradial amputee population. The ADL arm is functional open-source below-elbow prosthesis. This device is bodypowered; featuring a hand terminal device with thumb abduction and adduction, and wrist pronation and supination functionality. Elbow flexion of the residual limb is used to actuate the terminal device. The prosthesis requires no existing prosthetic hardware; and the majority of parts can be 3D printed. The ADL arm is designed to reliably perform the grasps required by the non-dominant hand in two-handed ADL activities. Device validation includes functional and simulated-use components. The functional assessment uses the Anthropomorphic Hand Assessment Protocol (AHAP); while the simulated-use assessment involves a practical ADL verification, and a usability assessment using healthy volunteer participants. The AHAP gives as result a grasping ability score (GAS) and partial GAS for ten grasp types associated with ADLs. The GAS represents the percentage of healthy limb function achievable by the prosthesis. The overall GAS of the ADL arm is found to be 68 %. The ADL arm achieved a partial GAS of greater than 75 % for four of five bimanual ADL grasps. A major design flaw resulted in a partial GAS of 33.3 % for the lateral pinch grasp type. The performance in this grasp, as well as others, would be greatly improved by the inclusion of a mechanism to lock the distal joint of the digits in extension during grasp. In this way, the hand would be better able to apply force to an object with the pads of the digits. Simulated-use validation of the ADL arm is performed on healthy participants using the designed bypass socket. The ADL assessment involves the completion of 86 ADL and instrumental ADL tasks; scored using the designed self-report questionnaire. The participant could perform all but seven tasks independently, and the perceived difficulty for tasks requiring the prosthesis was low overall. Seven healthy volunteers are used to assess the system usability. Participants performed a number of tasks and then completed the system usability scale (SUS). The perceived usability of the device is found to increase with increased device familiarity, yielding an overall score of 84.29. This result indicates that participants found the experience with the device to be ‘good' overall. In conclusion, the ADL arm is functionally competent and has proven its ability to assist in the performance of ADLs in a simulated-use environment; using healthy participants. A number of design modifications are recommended to overcome the limitations of the current design, which should be tested in the transradial amputee population to corroborate the results obtained in this study.
format Thesis
id oai:open.uct.ac.za:11427/36186
institution University of Cape Town (South Africa)
language eng
last_indexed 2026-06-10T12:32:31.718Z
license_str Not specified — see source repository
provenance_str_mv Harvested via OAI-PMH from UCTD — University of Cape Town Open Access Repository
publishDate 2022
publishDateRange 2022
publishDateSort 2022
publisher Department of Human Biology
publisherStr Department of Human Biology
record_format dspace
source_str UCTD — University of Cape Town Open Access Repository
spelling oai:open.uct.ac.za:11427/36186 Design and Development of an Open-Source ADL-Compliant Prosthetic Arm for Transradial Amputees Timm, Lara Sivarasu, Sudesh Biomedical Engineering Transradial amputation is traumatic – leading to the amputee having a limited ability to perform activities of daily living (ADLs). Below-elbow prostheses are prescribed. The high cost associated with prostheses results in many amputees in low-to-middleincome countries relying on government subsidised devices, which are cosmetic rather than functional, or none at all. Open-source prostheses have the potential to increase the accessibility of functional prosthetic arms, but at present are not optimised to assist the dominant hand in performing bimanual ADLs. The aim of this study is thus to design and experimentally validate an open-source prosthetic arm that is functionally optimised for the performance of ADLs in the unilateral transradial amputee population. The ADL arm is functional open-source below-elbow prosthesis. This device is bodypowered; featuring a hand terminal device with thumb abduction and adduction, and wrist pronation and supination functionality. Elbow flexion of the residual limb is used to actuate the terminal device. The prosthesis requires no existing prosthetic hardware; and the majority of parts can be 3D printed. The ADL arm is designed to reliably perform the grasps required by the non-dominant hand in two-handed ADL activities. Device validation includes functional and simulated-use components. The functional assessment uses the Anthropomorphic Hand Assessment Protocol (AHAP); while the simulated-use assessment involves a practical ADL verification, and a usability assessment using healthy volunteer participants. The AHAP gives as result a grasping ability score (GAS) and partial GAS for ten grasp types associated with ADLs. The GAS represents the percentage of healthy limb function achievable by the prosthesis. The overall GAS of the ADL arm is found to be 68 %. The ADL arm achieved a partial GAS of greater than 75 % for four of five bimanual ADL grasps. A major design flaw resulted in a partial GAS of 33.3 % for the lateral pinch grasp type. The performance in this grasp, as well as others, would be greatly improved by the inclusion of a mechanism to lock the distal joint of the digits in extension during grasp. In this way, the hand would be better able to apply force to an object with the pads of the digits. Simulated-use validation of the ADL arm is performed on healthy participants using the designed bypass socket. The ADL assessment involves the completion of 86 ADL and instrumental ADL tasks; scored using the designed self-report questionnaire. The participant could perform all but seven tasks independently, and the perceived difficulty for tasks requiring the prosthesis was low overall. Seven healthy volunteers are used to assess the system usability. Participants performed a number of tasks and then completed the system usability scale (SUS). The perceived usability of the device is found to increase with increased device familiarity, yielding an overall score of 84.29. This result indicates that participants found the experience with the device to be ‘good' overall. In conclusion, the ADL arm is functionally competent and has proven its ability to assist in the performance of ADLs in a simulated-use environment; using healthy participants. A number of design modifications are recommended to overcome the limitations of the current design, which should be tested in the transradial amputee population to corroborate the results obtained in this study. 2022-03-22T09:24:54Z 2022-03-22T09:24:54Z 2021 2022-03-22T05:58:02Z Master Thesis Masters MSc http://hdl.handle.net/11427/36186 eng application/pdf Department of Human Biology Faculty of Health Sciences
spellingShingle Biomedical Engineering
Timm, Lara
Design and Development of an Open-Source ADL-Compliant Prosthetic Arm for Transradial Amputees
thesis_degree_str Master's
title Design and Development of an Open-Source ADL-Compliant Prosthetic Arm for Transradial Amputees
title_full Design and Development of an Open-Source ADL-Compliant Prosthetic Arm for Transradial Amputees
title_fullStr Design and Development of an Open-Source ADL-Compliant Prosthetic Arm for Transradial Amputees
title_full_unstemmed Design and Development of an Open-Source ADL-Compliant Prosthetic Arm for Transradial Amputees
title_short Design and Development of an Open-Source ADL-Compliant Prosthetic Arm for Transradial Amputees
title_sort design and development of an open source adl compliant prosthetic arm for transradial amputees
topic Biomedical Engineering
url http://hdl.handle.net/11427/36186
work_keys_str_mv AT timmlara designanddevelopmentofanopensourceadlcompliantprostheticarmfortransradialamputees